Some of the functionalities will not work if javascript off. Please enable Javascript for it.
Official Website of Rajiv Gandhi Institute of Petroleum Technology, Jais, Amethi, India. / Facilities / Fuel Testing Facilities


Fuel Testing Facilities

Bomb Calorimeter
Make and Model

HAMCO6E

Specifications
  • Temperature Range15°C to 35°C, Temperature resolution +/- 0.001°C.
  • As per ASTM D 4809, D 240, D E711 , IS -1350- 1966, BS 1016 PART 5-1967 , IP 12/63 T.
  • CV Range 1000 to 20000 Cal /gm.
  • RSD +/- 1.0%.
  • CV Determination of Any Solid or Liquid.
Cloud and Pour Point Testing Apparatus
Make and Model

Specifications
  • Type of Chamber: Mechanically refrigerated cooled cascade type
  • Cooling Stage: Double stage with double compressor
  • Refrigerant: CFC free
  • Max working temperature: 55℃
  • Compartment: single
  • Insulation: Dense puff
  • Chamber cover: Made of PP with Bakelite knob having center hole
  • Mounting: Floor mounted
  • Movability: Mounted on four castor wheel with brakes on two front wheel for easy movability
Flash and Fire Point Tester
Make and Model

Abel

Specifications
  • Type- Manual Model
  • Measuring Range- 30°C to +110°C
  • Sample Cooling /Heating- Cooling-By Cooling Coil in the Bath, Heating-By Water immersion Heater
  • String- Manual
  • Temperature Sensor- Thermometer
  • Ignition Source- Gas ignition with automatic lighting or electric ignition.Interchageable
  • Power Supply- 100/120VAC or 220/240VAC, 50/60Hz
Redwood Viscometer
Make and Model

Specifications
  • Type of Bath :- Electrically Heated Immersion Heater
  • Max Working Temperature- 99℃
  • Continuous Working Temperature- 95℃
  • Capacity- 1 Sample at a time
  • Accuracy- +/- 3℃
  • Readability- +/- 3℃
  • Stirrer- Motorized Stirrer With Stirrer Blade
  • Top Lid made of Stainless Steel With Provision of mounting of one thermometer
Reid Vapor Pressure Apparatus
Make and Model

Specifications
  • Bath for double test with Micro-PID digital temperature controller
  • Vapor pressure bomb consisting of one air and one gasoline chamber without cock, Gasoline chamber with cock & Pressure Gauge supplied
  • Temperature range : Above ambient to +100°C
Smoke Point Detector
Make and Model

Specifications
  • Assessment of combustion quality of Kerosene and ATF